CHAPTER 2 - ESTABLISHING COMMUNICATION BETWEEN YOUR METER REMOTE AND YOUR PUMP
Changing the RF channel on your meter and pump
The RF channel on your meter remote must always match the RF channel on your pump. When you pair
your meter remote and pump, the RF channel is automatically set to match on both devices. To avoid
interference from another device or improve communication between your meter remote and pump,
you can also manually change/set the RF channel on your meter remote and pump to match.
NOTE: The "RF Channel" menu option on the RF set-up screen on your meter
remote is only available if your meter remote and pump are paired.
Manually set the RF channel on your meter remote
1
Go to RF Channel on your meter remote display
Press
on your meter remote to highlight "RF Channel" on the RF Set-up
screen. Then press
on your meter remote.
A Notification screen will appear on your meter remote display to remind you
to set the RF channel on your meter remote to match the one on your pump.
Press
to confirm the Notification on your meter remote display.
2
Set the RF Channel on your meter remote
Press
on your meter remote to manually select the desired channel.
Then press
on your meter remote.
After making your selection, you will return to the RF Set-up screen on your
meter remote display. You will then need to manually set the RF channel on
your pump to match the channel you set on your meter remote.
Manually set the RF channel on your pump
1
Go to Channel on your pump display
Go to the SETUP ADV 10 screen on your pump display. Press
until the "Channel" field is highlighted. Then press
highlight is flashing.
2
Set the RF Channel on your pump
Press
or
on your pump to change the current channel to match the RF
channel you selected on your meter remote. Then press
"Next" will be highlighted on your pump display. With "Home" highlighted,
press
on your pump to return to the Home screen on your pump display.
on your pump
on your pump so that the
on your pump.
186
RF Set-up
RF on/off
RF Channel
Pairing
RF Test
Meter remote display
Notification
Please use the same RF
channel as the paired
pump!
Confirm
Meter remote display
RF Channel
Selected RF Channel
1
p q
Meter remote display
SETUP ADV 10
METER
RF
ON
Search
OFF
Channel
9
[ABCD1234]
Home
Next
Pump display
SETUP ADV 10
METER
RF
ON
Search
OFF
Channel
1
[ABCD1234]
Home
Next
Pump display