CHAPTER 6 - TROUBLESHOOTING YOUR ONETOUCH
Notification
Unable to communicate
with pump!
Confirm
Notification
Function not available. All
communication to
connected RF devices is
stopped!
Confirm
Notification
Unable to communicate
with pump!
Bolus settings may
not be current.
Verify & edit bolus
settings as needed.
Confirm
Notification
Stored meter calculator
settings will be
overwritten by settings
from pump.
Confirm
®
INDICATES
Your meter remote and pump are
unable to communicate. Possible
causes are that the devices are not
within RF range or there is RF
interference. You will not be able
to use your meter remote to
deliver insulin from your pump
when the devices are unable to
communicate.
Your meter remote and pump are
unable to communicate. A possible
cause is that RF communication has
been deactivated.
You have accessed the bolus
calculator on either the ezCarb or
ezBG Bolus screens, but your
meter remote and pump are
unable to communicate. The
ezCarb and ezBG Bolus screens
will use the bolus calculator values
last set and saved on your pump.
Your meter remote and pump are
now paired, and the bolus
calculator settings from Calculator
Set-up will be replaced by those
last set and saved on your pump.
211
PING
SYSTEM
Press
to confirm the Notification.
See if your meter remote and pump
are within RF range (about 10 feet/
3.0 meters of one another), and
use the RF Test feature on your
meter remote to check the strength
and quality of the RF signal (see
pages 185–186). In cases where
you are not able to use your meter
remote to access pump functions,
you will still be able to deliver
insulin directly from your pump or
as an insulin injection.
Press
to confirm the Notification.
Check to see if you have
deactivated RF communication on
either your meter remote or pump
(see pages 182–184). In cases
where you are not able to use your
meter remote to access pump
functions, you will still be able to
deliver insulin directly from your
pump or as an insulin injection.
The values last set and saved on
your pump may not be appropriate
for the current time of day. Press
to confirm the Notification. You may
edit the values as necessary in
Calculator Set-up under the Meter
Settings screen if your meter
remote and pump are unpaired.
You may also edit the values
directly on the ezCarb and ezBG
Bolus screens.
Press
to confirm the Notification.
You may edit the settings on the
ezCarb and ezBG Bolus screens.
WHAT TO DO