Animas OneTouch Ping Owner's Booklet page 98

Glucose management system
Hide thumbs Also See for OneTouch Ping:
Table of Contents

Advertisement

CHAPTER 13 - SAFETY SYSTEM AND ALARMS
Alerts, Warnings and Alarms
Alarm: Call Service
Cause
Hardware or software problem detected.
Effect
All deliveries stop.
Message
Continuous until battery is removed.
Action
Press
be silenced once).
Call Customer Service at 1 877 937-7867.
Beeps/Vib
User-selected, every 3 minutes. If not confirmed,
progresses to sweep/vibe within one hour.
NOTE: Some Call Service Alarms have a unique sound/vibration sequence
and cannot be silenced by pressing
For these Alarms the usual progression is replaced by 3 chirps/vib repeated
every 9 minutes for the first half hour. This is followed by 4 long tones/vib
after that.
Alarm: Auto-Off
Cause
Suspended due to no key press in user-set time period.
Effect
All deliveries stop.
Message
Continuous until confirmed.
Action
Press
Advanced menu.
MAX volume every 3 minutes. If not confirmed,
Beeps/Vib
progresses to sweep/vibe within one hour.
(Once confirmed, No Prime warning triggered, see
No Prime Warning screen, page 81.)
to silence alarm for 30 minutes (alarm can only
.
to confirm. Adjust time period in the Setup
84
ALARM
CALL SERVICE
No delivery.
054-0108
Remove battery
to silence
the alarm.
ALARM
AUTO-OFF
No delivery.
No button
presses in last
XX hours.
Confirm

Advertisement