Animas OneTouch Ping Owner's Booklet page 92

Glucose management system
Hide thumbs Also See for OneTouch Ping:
Table of Contents

Advertisement

CHAPTER 13 - SAFETY SYSTEM AND ALARMS
Alerts, Warnings and Alarms
Warning: Basal Delivery Suspended
Cause
Basal Edit was not saved.
Effect
Basal delivery stopped.
Message
Displayed when manually awakened until confirmed.
Action
Press
select "Save/Review".
Beeps/Vib
User selected, every 3 minutes until confirmed. If not
confirmed, progresses to sweep/vibe within one hour.
Warning: Suspend
Cause
Pump suspended manually.
Effect
All deliveries stop.
Message
Displayed each time pump is awakened until confirmed
and once every 15 minutes until action is taken.
Action
Press
Beeps/Vib
User selected, once every 15 min. No progression if
confirmed each time displayed. Sweep/vibe within one
hour if not confirmed.
Warning: No Cartridge Detected, Deliveries Disabled
Cause
No cartridge detected after "Load cartridge" step during
Rewind/Prime.
Effect
No deliveries.
Message
Displayed when manually awakened until confirmed.
Action
Press
completed with cartridge properly in place.
Beeps/Vib
User selected, once every 3 min. No progression if
confirmed each time displayed. Sweep/vibe within one
hour if not confirmed.
to select "Edit Basal". Review basal edits and
to confirm. Resume delivery.
to confirm. Be sure Rewind/Prime sequence is
78
Warning
Basal edit
not saved.
Basal
delivery
suspended.
Edit Basal
Warning
No delivery.
Pump is
suspended.
Confirm
Warning
No cartridge
detected.
Delivery
disabled.
Confirm

Advertisement

Table of Contents
loading

Table of Contents