CHAPTER 3 - INTRODUCTION TO YOUR ONETOUCH
CHAPTER 3 - INTRODUCTION TO YOUR ONETOUCH
Contrast Button
Pressing this button adjusts the contrast of your display. There are three contrast levels: Dim, Default
and Bright. To preserve battery life, your pump display will Auto-dim when a button is not pressed for
half the time your display time-out is set. While in Auto-dim mode, you can restore the default contrast
level you have set by pressing the
Auto-dim mode will restore the default contrast level as well as perform the function of the key. To
adjust contrast during a Call Service alarm, you must use the
pages 49–58.
NOTE: When viewing your pump display in bright sunlight, it is recommended you shade the screen or
move to a shady area for best visibility.
Audio Bolus/ezBolus
This button allows you to program a bolus without looking at your pump, by using audible tones to
confirm programming and delivery. If you choose not to activate the Audio Bolus feature, this button
provides a shortcut to the Normal Bolus screen.
NOTE: When you first use the Audio Bolus feature, you should always look at the screen to confirm
correct programming until you are comfortable with using audio feedback to program a bolus. See
Chapter 10 in Section I, pages 49–58.
Battery Cap/Vent
This cap unscrews easily with a coin or the battery cap tool to replace and secure your battery. There is
an o-ring around the cap, which prevents water from entering. The battery cap also is equipped with a
built-in vent to allow air to enter your pump to maintain pressurization but prevent water from entering.
Be careful not to over tighten the battery cap. See Chapter 4 in Section I, pages 15–26.
Primary Vent
This vent is part of the redundant vent safety
system, which allows air inside your pump
to maintain equalized pressure but prevents
water from getting inside.
WARNING: Under no circumstances should you introduce any kind of sharp object into the
vent openings to clean them. Doing so will compromise your pump's waterproof capabilities. If
at any time you suspect the vent openings are clogged, replace the battery cap or call Animas
Customer Service for questions about the Primary Vent.
®
®
button on top of your pump. Pressing a function button while in
12
PING
INSULIN PUMP
PING
INSULIN PUMP
button. See Chapter 10 in Section I,
Cartridge Cap
Primary Vent
Battery Cap/Vent
®