CHAPTER 11 - USING ADVANCED FEATURES
ezBG
This feature allows you to enter your BG reading and your pump will automatically calculate a BG
correction bolus based on the ISF and BG Target range for the current time of day. If the IOB feature
is activated, your pump will calculate a reduced dose for high BG correction boluses.
1.
From the BOLUS MENU, select "ezBG". Press the
BOLUS MENU
Normal
ezCarb
ezBG
Combo Bolus
Reminders
Main Menu
2.
The "Actual" field will be highlighted and flashing
to indicate Edit mode. Use the
your actual BG reading. Press the
the entry and exit Edit mode.
3.
Check to be sure the BG Target and Insulin Sensitivity Factor (ISF) are
correct. Your health care team will give you these values. If you need to edit
these fields, scroll up to highlight the field and press
Use
/
buttons to change target. Press
ezBG
4.
mg/dL
Actual
224
Target
-
120
= +
104
ISF
37 mg/dL
Show Result
Main Menu
/
buttons to enter
button to confirm
to confirm and to exit Edit mode.
"Show Result" is highlighted. Press
65
button.
ezBG
mg/dL
Actual
224
Target
-
120
= +
104
ISF
37 mg/dL
Show Result
Main Menu
to activate Edit mode.
.
ezBG
mg/dL
Actual
224
Target
-
120
= +
104
ISF
37 mg/dL
Show Result
Main Menu
ezBG
mg/dL
Actual
224
Target
-
120
= +
104
ISF
37
mg/dL
Show Result
Main Menu