Animas OneTouch Ping Owner's Booklet page 214

Glucose management system
Hide thumbs Also See for OneTouch Ping:
Table of Contents

Advertisement

CHAPTER 4 - INSULIN DELIVERY FUNCTIONS ON YOUR METER REMOTE
4c
If all your entries are correct, press
If you need to adjust any entry, press
change and then press
NOTE: Changes made to your IS Factor or BG Target during ezCarb calculations apply to this bolus only
and will not affect the IS Factor or BG Target you have stored in your pump.
Pump Alert
LOW BG
Treat low BG.
No bolus recommended.
Monitor BG.
Confirm
Press
on your meter remote or pump to confirm the Alert. Treat a LOW BG
or HIGH BG immediately according to your health care professional's
recommendation.
5
Review and deliver your ezCarb Bolus
The Bolus Total screen shows the Carb and BG correction amounts
calculated from your previous entries. If the IOB feature is enabled on your
pump, the suggested total bolus amount will be adjusted accordingly.
"Total" units are rounded to the nearest .05 units. The amount field will
be highlighted and flashing, and will display 0.00 units.
Bolus Total
Carb
2.25U
BG
+
2.70U
IOB
– – . – –U
Total
=
4.95U
4.95
Go
U
Type Normal
Main Menu
5b
"Go" is highlighted. You can deliver the ezCarb Bolus either as a Normal
Bolus or as a Combo Bolus by making your selection in the "Type" field. The
pre-set delivery type for an ezCarb Bolus is Normal.
with "Show Result" highlighted on the BG Correct screen.
to first highlight it, and then press
again.
Before calculating an estimated BG correction, your meter remote will first
check to see if your Actual BG is within the range 70–250 mg/dL. If your
Actual BG falls below 70 mg/dL or above 250 mg/dL, you will be prompted
with either a LOW BG or HIGH BG Pump Alert screen.
NOTE: If you have not enabled the IOB feature on your pump, a set of
dashes ("––.––") will appear as the IOB amount.
5a
Press
to enter either the suggested "Total" units or a different bolus
amount. Press
once you have selected the desired bolus amount.
196
. Use
to make the
Pump Alert
HIGH BG
Treat high BG.
Check site.
Check ketones.
Monitor BG.
Confirm
Bolus Total
Carb
2.25U
BG
+
2.70U
IOB
––. – – U
Total
=
4.95U
0.00
Go
U
Type Normal
Main Menu
Bolus Total
Carb
2.25U
BG
+
2.70U
IOB
–– . –– U
Total
=
4.95U
4.95
Go
U
Type Normal
Main Menu