Animas OneTouch Ping Owner's Booklet page 115

Glucose management system
Hide thumbs Also See for OneTouch Ping:
Table of Contents

Advertisement

CHAPTER 18 - WARRANTY AND OTHER TECHNICAL INFORMATION
Flow Rate Accuracy
Delivery Mode
Bolus
Basal
Occlusion Detection time to alarm*
Bolus 3 units or more
0.025U/Hr basal
1.0U/Hr basal
*Certain factors, such as the presence of air in the infusion set or the cartridge and/or ambient
temperature changes, can delay an occlusion alarm.
Occlusion Pressure Threshold:
Bolus Volume after Occlusion release: • 1.0U max with occlusion sensitivity set to high
(1.0U/Hr basal)
Delivery Rates:
Insulin Types Used:
Basal Rate Range:
Bolus Range:
Protection from equipment error:
Continuous Operation, Internally Powered Device
Type BF Medical Equipment (Patient isolated, not defibrillator protected)
Watertight Equipment, IPX8 (protected against the effects of submersion, tested at 12 feet (3.6 meters)
for 24 hours).
Infrared communication port
Accuracy
+/- 5%
+/- 5%
Typical
8 sec.
72 hrs.
90 minutes
75 kPa typical, 241 kPa max.
• 3.0U max with occlusion sensitivity set to low
• Bolus, under 1U: 1.1 to 2.2U/sec
• Bolus, 1U or more (normal delivery speed): 0.5 to 0.9U/sec
• Bolus, 1U or more (slow delivery speed): 0.2 to 0.4U/sec
• Prime: 1.8 to 2.9U/sec
Rapid-acting U100 insulin or regular (short-acting) U100
insulin
0.025-25U/Hr in 0.025U/Hr steps
0.05-35U in 0.05U steps
More than 1.5 million redundant safety cross-checks per day
for both hardware and software functionality.
101
Maximum
30 sec.
120 hrs.
3 hrs.