CHAPTER 8 - CONTROL SOLUTION TESTING

Performing a control solution test

1
Check the code on the test strip vial before inserting the test strip.
2
Insert a test strip to turn on your meter remote.
Make sure the three contact bars are facing you. Push the test strip in as far as
it will go. Do Not bend the test strip.
An all-black start-up screen will be followed by an
hourglass symbol and then the Match Code screen.
CAUTION: If the graphics appear to be different, call Customer Service at 1 877 937-7867.
There may be a problem with your meter remote.
3
Match the code displayed on your meter remote with the code on the test strip vial.
If the code on your meter remote does not match the code on the test strip
vial, press
to match the code number on the test strip vial. The new code
number will flash on the display for three seconds, after which the display will
advance to the Test/Apply Blood screen.
If the codes already match, press
change after three seconds, the display will advance to the Test/Apply Blood screen.
4
Set your meter remote for a control solution test.
Press
to change "Apply Blood" to "Control
Solution". To mark a test as a control solution
test, you must change "Apply Blood" to "Control
Solution" before you apply the solution. It cannot
be changed later.
to go to the Test/Apply Blood screen. When you do not make a
Test
Apply Blood
160
Test
Match Code
12
(Example)
Test
Match Code
12
Test
Control Solution
p q
p q
p q
p q