Installation

Cartridge Pack Rinse-Up

7. Close door.
8. DO NOT install the 0.2 micron filter and bell assembly at this time.
Cartridge Pack Rinse-Up
1. If not already installed, install the hose barb into the fitting in the dispense valve.
2. Turn unit on STANDBY. The RO system will automatically turn ON, flush the RO membrane
for 30 seconds, and then fill the reservoir. The reservoir needs at least three liters of RO water to
perform the flush.
3. With at least three liters of water in the reservoir, turn unit on and from STANDBY press the
START/STOP button to begin unit operation. NOTE: It may take another minute or two for
the water to fill the new cartridge pack and begin flowing out the dispense valve.
4. Rinse approximately three liters of water through the dispense valve to drain.
5. Close the dispense valve.
6. Remove the hose barb from the dispense valve and proceed to the 0.2 Micron Final Filter
Replacement section of this manual. Retain hose barb for future use.
7. Place unit in RUN mode until desired purity is achieved (>15 megohm-cm).

Control Panel Wall Mount Bracket

Figure 3-24 Control Panel Mounted in Remote Location
For your convenience, the control panel can be removed from the unit and mounted at a convenient
location within 10 ft. (3.1 m.) of the unit. To remove the control panel from the unit:
1. Turn the unit OFF and disconnect it from the power supply.
2. Slide the control panel upward about 1/4" (0.64cm) to 3/8" (0.95 cm) just enough to clear
connector on top cover, pull forward, and remove. The control panel fits tightly in the unit so it
may be difficult to remove.
3. Replace the control panel with the blank panel provided with the unit to protect the electrical
connector as shown in Figure 3-25.
4. Using the display wall bracket as a template, locate and drill the mounting holes in the wall. A
minimum of two (customer-supplied) fasteners will be required. Attach the wall bracket to the
wall using the customer-supplied fasteners as shown in Figure 3-26.