Locating Unit; Bench Mounting; Wall Mounting - Thermo Scientific 1403 Series Operating Manual

Barnstead tii water purification system series 1403 and 1404
Table of Contents

Advertisement

Locating Unit

Locate storage reservoir approximately 2 - 5 ft. (1.52 m) from the Thermo Scientific TII for electrical
connections.
Refer to the Thermo Scientific TII storage reservoir manual or customer-supplied reservoir manual for
mounting instructions. The outlet of the Thermo Scientific TII storage reservoir must be above or at
the same level as the inlet of the Thermo Scientific TII.

Bench Mounting

Place the Thermo Scientific TII on a bench top that is accessible to electricity and an atmospherically
vented drain.

Wall mounting

Wall composition, condition and construction, as well as fastener type, must be considered when
mounting this unit. The mounting surface and fasteners selected must be capable of supporting a
minimum of 150 lbs. (68kg). Inadequate support and/or fasteners may result in damage to mounting
surface and/or equipment. If you are unsure of mounting surface composition, condition and
construction or correct fasteners, consult your building maintenance group or contractor. Refer to
Section
Install the Thermo Scientific TII on a wall in a convenient location that is accessible to an
atmospherically vented drain and electricity.
1. Locate the wall bracket packed separately from the unit.
2. Using the wall bracket as a template, locate and drill the mounting holes in the wall. A minimum
of four (customer-supplied) fasteners will be required — two on the top and two on the bottom.
3. Attach the wall bracket to the wall using customer-supplied fasteners.
4. Remove the locking screws on each side of the wall bracket.
5. Pull the two locking slides on each side of the wall bracket out as far as they will go.
6. Hang the unit on the wall bracket by sliding the mounting pins into the wall bracket slots.
7. Push the locking slides on each side of the wall bracket in as far as they will go.
8. Replace the locking screws.
Thermo Scientific
Technical
Specifications, for clearance requirements.
WARNING Do not locate the Thermo Scientific TII directly over equipment that
requires electrical service. Routine maintenance of this unit may involve water spillage and
subsequent electrical shock hazard if improperly located.
Note The outlet of Thermo Scientific TII storage reservoir must be above or at the same
level as the inlet of the Thermo Scientific TII.
Installation
Bench Mounting
TII Water System
3-3