Reservoir Connection - Thermo Scientific 1403 Series Operating Manual

Barnstead tii water purification system series 1403 and 1404
Table of Contents

Advertisement

Installation

Reservoir Connection

4. Turn on power to unit.
5. Ensure the unit is in IDLE mode.
6. Press MENU and use the arrows to scroll to "RO Options", press OK. Scroll to "Membrane
Rinse" and press OK.
7. Follow the prompts on the screen to rinse membranes. The prompts will ensure that you do not
have a tank connected to the TII unit by asking if the tank is empty and if the tubing from the
tank goes to drain. Answer yes to both questions. Once initiated, the rinse procedure will last
two hours, and will stop automatically.
Reservoir Connection
After performing an RO membrane rinse, connect the TII to a storage reservoir. Refer to the Thermo
Scientific TII Storage Reservoir manual.
Figure 3-20 Thermo Scientific TII and Thermo Scientific TII Reservoir System Set-Up
3-14
TII Water System
CAUTION Do not allow the Thermo Scientific TII to operate unless water is available to
the unit.
WARNING Use a properly grounded electrical outlet of correct voltage and current
handling capacity.
Note Install reservoir at the same height or higher than the TII system. Outlet of the
reservoir must be at or above the inlet of the TII.
Thermo Scientific