Push-To-Connect Fitting Tubing Install; Push-To-Connect Tubing Removal - Thermo Scientific 1403 Series Operating Manual

Barnstead tii water purification system series 1403 and 1404
Table of Contents

Advertisement

Installation

Push-To-Connect Tubing Removal

Push-to-Connect Fitting Tubing Install

The following instructions will apply when you need to attach a piece of tubing to your Thermo
Scientific TII during installation, unless otherwise noted in the installation instructions. To make
tubing connections:
1. Make sure the tubing is cut off reasonably square and that no plastic burrs or ridges are present.
2. Mark from end of tube an insertion length of 3/4" (1.9 cm).
3. Wet the tube end with water and insert the tube straight into the fitting until it bottoms out on
the interior shoulder and the insertion mark is no longer visible.
Figure 3-7 Push-To-Connect Fitting Tubing Installation
Push-To-Connect Tubing Removal
1. Using the tool provided (AYX23), push the collet toward the body while pulling on the tubing to
release the tube, as shown in Figure 3-8.
Figure 3-8 Push-To-Connect Tubing Removal