RO Membrane Rinse

All the RO prefilters and RO membranes and drain tubing should already be installed at this point. If
they have not, refer to
Preparation
When new RO membranes are installed in the Thermo Scientific TII, the membranes should be rinsed
to remove all traces of preservative from the membranes.
1. Ensure that the feed water and drain connections are in place and power connections are made.
Refer to connection information in this section if needed.
2. Verify atmospheric drain tubing is attached.
3. Connect tubing from "FROM TANK" on the TII unit connection to an atmospheric drain.
Product water (not usable) will flow from this tube during the RO membrane rinse procedure.
Thermo Scientific
on
page 3-4
for more explicit details.