Computer/Printer Setup
8. Open procomm executable.
9. Press ALT-F10 for configuration screen.
10. Enter into Line Settings screen by typing ALT-P.
11. Type in 11 <Enter>, sets parameters as 9600, 8, N, 1.
12. Type in 21 <Enter>, sets to (COM2:) or (Type in 20 <Enter>, sets to (COM1:), depending on
your computer.
13. Type in 24 <Enter>, saves settings.
14. Press ESC to exit setup menu.
15. Procomm is now ready to accept input from the Thermo Scientific TII.
16. To begin downloading a file, Type ALT-F1, and give the file an appropriate name when
prompted.
The optional printer (part no. AY1137X1) provides a paper tape record for future reference. During
normal recirculation, the printer prints date, time, purity and temperature readings once every
minute.

Connecting and Starting the Printer

1. Make sure the printer and Thermo Scientific TII are turned OFF.
2. Connect the Thermo Scientific TII to the printer via the RS232 port on the back left side of the
unit. Use the tan printer cable included with the Epson printer. Plug the printer power cord into
an electrical outlet.
3. Turn power on to the Thermo Scientific TII and printer.
4. Press MENU and use the up or down arrows until display reads, "SYSTEM OPTIONS."
5. Press OK.
6. From the "Printer Attached" display, press the up or down arrow for "YES".
7. Press OK.
8. Press the BACK arrow twice to return to the STANDBY mode.
9. During normal recirculation mode, a purity and temperature reading will be printed once every
12 minutes.