Thermo Scientific 1403 Series Operating Manual page 59

Barnstead tii water purification system series 1403 and 1404
Table of Contents

Advertisement

Message
Alt Res Float Err
Purity Meter Error
Meter Check Failed
DI Temp Error
Low DI
Thermo Scientific
Possible Causes
Float not operating correctly.
Floats not wired correctly.
Floats not installed correctly.
Alternate reservoir in use.
Float cable not securely connected.
Float not operating correctly.
Floats not wired correctly.
Floats not installed correctly.
Thermo Scientific TII reservoir in
use.
Control board out of tolerance.
Meter on control board out of cali-
bration.
Meter on control board out of toler-
ance.
Resistivity cell not connected to
board.
Resistivity cell temperature sensor
out of tolerance.
Reservoir water temperature outside
of allowable range.
Control board out of tolerance.
Check cell tolerance.
Cartridge pack near exhaustion.
Cartridge pack not in place (saniti-
zation block installed?).
DI cell out of tolerance.
DI meter on control board out of
tolerance.
Check cell tolerance
Maintenance and Servicing
System Alarms and Errors
Action
Check float operation.
Check float wiring.
Check float installation.
Using Thermo Scientific TII menu
screens, set reservoir type to "Alter-
nate" (in "System Options").
Check float cable connection.
Check float operation.
Check float wiring.
Check float installation.
Using Thermo Scientific TII menu
screens, set reservoir type to
"Thermo Scientific TII" (in "System
Options"). Ensure interface connec-
tor is securely connected to the
Thermo Scientific TII and the
Thermo Scientific TII storage reser-
voir.
Replace control board (PC1403X3).
Calibrate meter using NIST mod-
ule.
Replace control board (PC1403X3).
Follow the enclosed calibration
instructions.
Check resistivity cell lead connec-
tion on board.
Replace resistivity cell (E1403X6).
Check the temperature of the stor-
age reservoir water.
Replace control board (PC1403X3).
Ensure that reading on tag agrees
with reading in software.
Install a new cartridge pack
(D502137) and reset timer.
Check that cartridge pack is
installed.
Replace resistivity cell (E1403X6).
Replace control board (PC1403X3).
Ensure that reading on tag agrees
with reading in software.
TII Water System
5-9