To measure average TX Power when a mobile is in the Connect state and a PCS
Interface is configured and enabled, use the following measurement procedure.
Average TX Power with PCS Interface measurement procedure
1. Set up the PCS Interface as directed in the sections under
Adapter" on page 32
2. Establish a call between the mobile and the Test System on a digital traffic channel
(DTC) (or an analog voice channel (AVC) if you prefer). When established, the
Connect indicator on the CALL CONTROL screen is 'lit' and the Traffic Chan
Assignment - Type: indicates DTC (or AVC). See
3. Select Meas from the Data/Meas field.
4. Average TX Power is displayed, along with other relevant analog measurements.
Measuring TX Power on a Digital Traffic Channel (DTC) with a PCS Interface
.
To go to the
ANALOG MEAS
screen.
"About the Cellular
"Registration" on page 93
Confirms
DTC
PCS_POWER.tif
.