Rear Panel Connectors
IQ RF OUT
252
S:\HP83206A\USRGUIDE\MANUAL\connect.fm
This rear-panel port outputs the carrier applied to the rear-panel CW RF IN
connector.
This port is normally connected to the Test Set's rear-panel IQ RF IN connector.
Operating Considerations
Even when the Digital Generator is not actively sending a message, the IQ
Modulator still receives symbol data.
Output impedance = 50
Nominal output level:
9.5 dBm ( 1.5 dB) when IQ modulated.
0 dBm un-modulated (CW).
Frequency range:
500 to 1000 MHz without IQ modulation (may have AM or FM from the Test Set).
810 to 956 MHz with IQ modulation.