Zeroing the
TDMA Average Power can be measured if your Test Set is equipped with
Average Power
Option 006, or if you have a PCS Interface connected. If you are not using the
Meter
PCS Interface, you need to zero the average power detector in the Test Set first
using the following procedure.
See
more information.
1. Disconnect all cables from the RF IN/OUT port of the Test Set.
2. Select CALL CNTL from the To Screen. (Use MORE to display the field CALL
3. Select DCCH in the System Type field.
4. Select CALL CNFG from the To Screen.
5. Select Zero from the Power Meter field. This zeros the power for the Test System.
6. Select CALL CNTL from the To Screen. This returns the Test Set to the primary call
Figure 3
Zeroing the Average Power Meter
"Average Power" on page 114
CNTL.)
When the Test Set is finished zeroing, the Zero field is no longer highlighted.
processing screen.
CALL CNTL
More
System Type
Power Meter
Chapter 1, Getting Started
and
"Power Measurements" on page 130
CALL CNFG
Get ready to test
for
CALL CNTL
35