HP 83206A User Manual page 141

Cellular adapter
Table of Contents

Advertisement

Measuring Relative
The following measurement procedure is for mobiles in the Connect state. Use the
TX Power on a
DIGITAL MEASUREMNTS screen to make measurements with the mobile
DTC
transmitting in the test mode (see
118).
1. Access the CALL CONTROL screen.
2. Establish a call between the mobile and the Test System on a digital traffic channel.
3. Select MEAS from the Data/Meas field.
4. Select EVM 1 from the Choices menu.
5. Relative TX Power is displayed, along with relevant digital modulation measurements.
Read
Relative
TX Power
Figure 55
Measuring TX Power on a Digital Traffic Channel (DTC)
When established, the Connect indicator on the CALL CONTROL screen is 'lit' and
the Traffic Chan Assignment - Type: indicates DTC. See
page 93
.
Chapter 4, Making Measurements
"Test mode measurement procedure" on page
To go to the DIGITAL
MEASUREMENTS
screen.
TX Power
"Registration" on
Confirms
DTC
Reltxpwr.tif
141