Chapter 4, Making Measurements

Adjacent Channel Power

Adjacent Channel Power
Figure 35
112
S:\HP83206A\USRGUIDE\MANUAL\meas.fm
Adjacent Channel Power is measured in either Connect state or test mode.The
measurement procedure described below is for the Connect state. To measure in
the test mode, see
"Test mode measurement procedure" on page
Adjacent Channel Power measures the power on six adjacent channels. These are
the three nearest channels both below and above the designated channel.
Adjacent Channel Power is measured on each adjacent channel relative to the
current channel.
Adjacent Channel Power measurement procedure
1. Make sure the Test Set is in Connect state. See
2. Select Meas from the Data/Meas field.
3. A field appears below the Data/Meas field. Select this field.
4. Select Adj Ch Pwr from the list of choices. The Adjacent Channel Power will be
displayed. This measurement can take several seconds to complete.
Adjacent Channel Power Display
118.
"Registration" on page 93
adjchpwr.tif
.