HP 83206A User Manual page 46

Cellular adapter
Table of Contents

Advertisement

Chapter 1, Getting Started
About the PCS Interface
Select the type of
mobile your are
testing: US PCS,
Cellular, or Band 10
(downband).
Figure 9
46
S:\HP83206A\USRGUIDE\MANUAL\getstart.fm
Specifying the Type of Mobile Station You Are Testing:
Even though you may be operating a PCS-compatible phone in the cellular band,
it still uses the PCS protocol to perform call processing. You can also test
non-PCS phones through the PCS Interface, so you still need to tell the test set
what type of mobile you are testing. This is done using the
DCCH CALL CONFIGURE II screen. See
To specify the type of mobile you are testing:
1. Access the CALL CONTROL screen by highlighting the More field and selecting
CALL CNTL from the list of choices.
2. Set the System Type to DCCH.
3. Select CALL CFG2 from the To Screen list.
4. Set the MS Capab field to indicate the type of mobile you are testing.
Select US PCS when testing PCS-capable mobiles (whether they are being tested
in the PCS band or not).
Select Cellular when testing standard cellular band TIA/EIA-136 mobiles.
Setting MS Capability and selecting Temperature Compensation.
MS Capab
Figure
9.
Periodically select the
Temperature
(Compensate) field to
compensate for
temperature changes
during operation.
Calconf2.tif