EVM is measured in either Connect state or test mode. The measurement
procedure described below is for the Connect state. To measure in the test mode,
see
Error Vector Magnitude (EVM) measures the magnitude (rms) of the error vector
between the measured and the ideal signal. EVM 1 measures over one burst. EVM
10 is the first 10 symbols of 10 bursts, averaged. Peak EVM is the peak value
measured over the burst.
The procedure below describes the measurement for EVM 1. EVM 10 is listed
below EVM 1 in the list of choices.
EVM measurement procedure
1. Make sure the Test Set is in Connect state. See
2. Select Meas from the Data/Meas field.
3. EVM 1 is displayed as default. If EVM is not displayed, continue with the next step.
4. A field appears below the Data/Meas field. Select this field.
5. Select EVM 1 from the list of choices. The measurement will display.
Figure 40
EVM Display