Performance Verification 5-8
3155 outputs a gated signal. Leave the external
pulse generator connected to the Model 3155 for
the next test.
Burst
1.
Configure the Model 3155 as follows:
Function:
Frequency:
Operating Mode:
Remote Commands:
2.
Set the oscilloscope and verify that the Model
3155 outputs a burst of 10 complete output
waveforms. Remove the external pulse generator
connection from the Model 3155 for the next test.
3155 User Manual
Sine wave
1.000 kHz
Burst
:RES
:OUTP ON
:FREQ 10e3
:INIT:CONT OFF
:BURST:STAT ON
:TRIG:COUN 10