Accuracy Specifications: 0.01% of reading to 50.00 MHz
Equipment: Counter
1.
Configure the Model 3155 as follows:
Function:
Frequency:
Amplitude:
Remote Commands: :RES
2.
Set the counter to frequency measurement.
3.
Connect the Model 3155 output to the counter's input.
Change the frequency setting as required for the test
and verify the reading on the counter display as follows:
3155 SETTING
10.00000 Hz
100.0000 Hz
1.000000 kHz
10.00000 kHz
100.0000 kHz
1.000000 kHz
10.00000 MHz
50.00000 MHz
Accuracy Specifications (1 kHz):
± (1% of reading +1 mV), to 799.9 mV
± (1% of reading +10 mV), to 8 V
Equipment: DMM, 50O feedthrough termination
1.
Configure the Model 3155 as follows:
Function:
Frequency:
Amplitude:
Remote Commands: :RES
Square
10 Hz
1 V
:OUTP ON
:FUNC:SHAP SQU
:FREQ 10
:VOLT 1
COUNTER READING
9.999000 Hz - 10.00100 Hz
99.99000 Hz - 100.0100 Hz
999.9000 Hz - 1.000100 kHz
9.999000 kHz - 10.00100 kHz
99.99000 kHz - 100.0100 kHz
999.9000 kHz - 1.000100 kHz
9.999000 MHz - 10.00100 MHz
49.99500 MHz - 50.00500 MHz
Sine
1.000 kHz
7.5 V
:OUTP ON
:FUNC:SHAP SIN
:FREQ 1000
:VOLT 15