Racal Instruments 3155 User Manual page 163

100ms/s arbitrary waveform generator
Table of Contents

Advertisement

User Manual 3155
ARM:SLOPe{POSitive|NEGative}
Purpose
This command sets the edge sensitivity of which the 3155 will stop in arm mode. The stop and start com-
mands are applied at the rear-panel TRIG IN connector.
POSitive sets the positive edge
NEGative sets the negative edge
Parameter type
Discrete
ARM:SLOPe?
Response and default
The 3155 returns POS, or NEG depending on the present setting. Default value is POS.
ARM:BREakpoint:POSition<position>
Purpose
This command programs the breakpoint position for the arm function.
<position> sets the breakpoint position
Parameter type
Numeric (integer only)
Parameter range
<position>
0 to "n" in units of waveform points, where "n" is the size of the active segment. The
breakpoint position can be programmed in increments of 4 points minimum.
ARM:BREakpoint:POSition?
Response and default
The 3155 returns the present breakpoint value. Default value is 0.
INITiate:CONTinuous{OFF|ON|0|1}
Purpose
This command sets the output to continuous operation or triggered operation. The trigger subsystem com-
mands affect the 3155 only after it is set to triggered operation by turning INIT:CONT OFF.
• ON or 1 sets the 3155 to continuous mode
OFF or 0 sets the 3155 to triggered mode
Parameter type
Discrete
INITiate:CONTinuous?
Response to query version
The 3155 returns 1 if the output is triggered, or 0 if the output is continuous. Default value is 0.