Dc Offset Characteristics - Racal Instruments 3155 User Manual

100ms/s arbitrary waveform generator
Table of Contents

Advertisement

DC Offset Charac-
teristics
Performance Verification 5-4
2.
Set the DMM to ACV measurements (RMS).
3.
Connect the Model 3155 output to DMM input. Termi-
nate the output with a 50O feedthrough termination. Set
the amplitude and verify DMM reading as follows:
3155 SETTING
7.50 V
750.0 mV
Accuracy Specifications:
± (1% of reading +1% of amplitude +200µV), ± 400mV window
± (1% of reading +1% of amplitude +2mV), ±4V window
Equipment: DMM, 50O feedthrough termination
1.
Configure the Model 3155 as follows:
Function:
Frequency:
Amplitude:
Remote Commands: :RES
2.
Set the DMM to DCV measurements
3.
Connect the Model 3155 output to the DMM input. Ter-
minate the output with a 50O feedthrough termination.
Set the Model 3155 offset and verify the DMM reading
as follows:
OFFS SETTING
±2.500 V
4.
Change the Model 3155 AMPL setting to 100 mV. Set
VOLT:OFFS and verify the DMM reading as follows:
OFFS SETTING
±250.0 mV
3155 User Manual
DMM READING
2.682 V – 2.622 V
268.2 mV – 262.2 mV
Sine
1.000 MHz
1 V
:OUTP ON
:FUNC SIN
:FREQ 1e6
:VOLT 1
:VOLT:OFFS 2.5
DMM READING
±2.537 V to ±2.463 V
DMM READING
±253.7 mV to ± 246.3 mV