Performance Verification; What's In This Chapter; Performance Checks; Environmental Conditions - Racal Instruments 3155 User Manual

100ms/s arbitrary waveform generator
Table of Contents

Advertisement

3155 User Manual
What's in This
Chapter
Performance
Checks
CAU TION
SENSITIVE ELECTRONIC DEVICES
DO NOT SHIP OR STORE NEAR
STRONG ELECTROSTATIC,
ELECTROMAGNETIC, MAGNETIC OR
RADIOACTIVE FIELDS
Environmental
Conditions

PERFORMANCE VERIFICATION

This chapter provides performance tests for the Model 3155
Waveform Synthesizer.
ALWAYS PERFORM DISASSEMBLY, REPAIR AND
CLEANING AT A STATIC SAFE WORKSTATION.
The following performance checks verify proper operation of
the instrument and should normally be used:
1.
As a part of the incoming inspection of the instrument
specifications;
2.
As part of the troubleshooting procedure;
3.
After any repair or adjustment before returning the i n -
strument to regular service.
Tests should be performed under laboratory conditions having
an ambient temperature of 25
of less than 80%. If the instrument has been subjected to con-
ditions outside these ranges, allow at least one additional hour
for the instrument to stabilize before beginning the adjustment
procedure. Always perform a self-test sequence before com-
mencing with the performance checks. The self-test, if exe-
cuted without any failure, ensures proper operation of the syn-
thesizer. If self-test failures have been encountered, the in-
strument first needs to be serviced and the source of failure
removed.
Chapter 5
CAUTION
C ±5
o
o
C and at relative humidity
Performance Verification 5-1