4
Triggering
FlexRay Triggering
N O T E
108
It is important to specify the correct bus because this setting affects CRC error
detection.
8 Press Auto Setup to perform the following actions:
•
Set the selected source channel's impedance to 50 Ohms.
•
Set the selected source channel's probe attenuation to 10:1.
•
Set the trigger level (on the selected source channel) to -300 mV.
•
Turns on trigger Noise Reject.
•
Turns on Serial Decode.
•
Sets the trigger type to FlexRay.
To set up the oscilloscope to capture a FlexRay signal, see
Signals"
on page 132.
After you have set up the oscilloscope to capture a FlexRay signal, you can then set up
triggers on frames, errors, or events.
To display FlexRay serial decode, see
Triggering on FlexRay Frames
1 In the front panel Trigger section, press [More].
2 In the More Trigger Menu, press Trigger; then, select the FlexRay trigger type.
3 Press Settings.
4 In the FlexRay Trigger Menu, press Trigger; then, select Frame.
"Setup for MIL-STD 1553
"FlexRay Serial Decode"
InfiniiVision Oscilloscopes User's Guide
on page 302.