Agilent Technologies InfiniiVision 5000 Series User Manual page 110

Infiniivision 5000/6000/7000 series oscilloscopes
Hide thumbs Also See for InfiniiVision 5000 Series:
Table of Contents

Advertisement

4
Triggering
N O T E
110
Triggering on FlexRay Errors
1 In the front panel Trigger section, press [More].
2 In the More Trigger Menu, press Trigger; then, select the FlexRay trigger type.
3 Press Settings.
4 In the FlexRay Trigger Menu, press Trigger; then, select Error.
5 Press the Errors softkey; then, select the error type:
All Errors
Header CRC Error — cyclic redundancy check error in the header.
Frame CRC Error — cyclic redundancy check error in the frame.
Since FlexRay errors occur infrequently it may be helpful to set the oscilloscope to
Normal trigger mode instead of Auto trigger mode. This prevents the oscilloscope from
Auto triggering while waiting for an error to occur. You may need to adjust trigger holdoff
to see a particular error when multiple errors exist.
Triggering on FlexRay Events
1 In the front panel Trigger section, press [More].
2 In the More Trigger Menu, press Trigger; then, select the FlexRay trigger type.
3 Press Settings.
InfiniiVision Oscilloscopes User's Guide