InfiniiVision Oscilloscopes User's Guide
Stop Condition — The oscilloscope triggers when data (SDA) transitions from low
to high while the clock (SCL) is high.
SDA
SCL
Address
Start
Condition
Missing Acknowledge — The oscilloscope triggers when SDA data is high during
any Ack SCL clock bit.
Address with no Ack — The oscilloscope triggers when the acknowledge of the
selected address field is false. The R/W bit is ignored.
Restart — The oscilloscope triggers when another start condition occurs before a
stop condition.
EEPROM Data Read — The trigger looks for EEPROM control byte value 1010xxx
on the SDA line, followed by a Read bit and an Ack bit. It then looks for the data
value and qualifier set by the Data softkey and the Data is softkey. When this event
occurs, the oscilloscope will trigger on the clock edge for the Ack bit after the data
byte. This data byte does not need to occur directly after the control byte.
SDA
SCL
Control byte
Start or
Restart Condition
Ack
Data
R/
Read
Data
Ack
R/
Triggering
Ack Stop
Condition
Ack
Trigger point
4
117