10
Serial Decode/Lister
294
4 Press the [Single] key (in the Run Control group on the front panel) to stop the
acquisition.
Pressing [Single] instead of [Stop] fills the maximum memory depth.
When zoomed out and viewing a large number of packets, the Lister may not be able
to display information for all packets. However, when you press the [Single] key the
Lister will contain all on-screen serial decode information.
5 Press the Scroll Lister softkey and use the Entry knob to scroll through the data.
Time tags in the Time column indicate the event time relative to the trigger point. The
time tags of events that are shown in the waveform display area are displayed with a
dark background.
6 Press the Zoom to Selection softkey (or push the Entry knob on 7000 Series
oscilloscopes) to center the waveform display at the time associated with the selected
Lister row and automatically set the horizontal scale setting.
7 Press the Undo Zoom softkey to return to the horizontal scale and delay settings
before the last Zoom to Selection.
8 Press the Options softkey to open the Lister Options Menu. In this menu, you can:
Enable or disable the Track Time option. When enabled, as you select different
Lister rows (using the Entry knob while acquisitions are stopped), the horizontal
delay changes to the Time of the selected row.