15
Reference
416
File Header
Number of Waveforms = N = Number of Segments
12 bytes
Number of Waveform Buffers = 1
Waveform Header 1
Index = 1
140 bytes
Time Tag = 0.0
Waveform Data
Buffer Type = 1 (floating point)
Header 1
Bytes per Point = 4
12 bytes
Voltage Data 1
buffer size
Number of Waveform Buffers = 1
Waveform Header 2
Index = 2
140 bytes
Time Tag = time between segment 1 and 2
Waveform Data
Buffer Type = 1 (floating point)
Header 2
Bytes per Point = 4
12 bytes
Voltage Data 2
buffer size
Number of Waveform Buffers = 1
Waveform Header N
Index = N
140 bytes
Time Tag = time between segment 1 and N
Waveform Data
Buffer Type = 1 (floating point)
Header N
Bytes per Point = 4
12 bytes
Voltage Data N
buffer size