Mil-Std 1553 Serial Decode - Agilent Technologies InfiniiVision 5000 Series User Manual

Infiniivision 5000/6000/7000 series oscilloscopes
Hide thumbs Also See for InfiniiVision 5000 Series:
Table of Contents

Advertisement

10
Serial Decode/Lister

MIL-STD 1553 Serial Decode

N O T E
320
To probe MIL-STD 1553 signals, see
To set up the oscilloscope to capture MIL-STD 1553 signals, see
1553 Signals"
on page 132.
Before you can select serial decode of MIL-STD 1553 signals, the trigger type must be set
to MIL-STD-1553 or one of the non-serial triggering modes. See
Triggering"
on page 133.
When MIL-STD-1553 is selected as the trigger mode, only MIL-STD-1553 decode is
available.
To set up MIL-STD 1553 serial decode:
1 Select the MIL-STD 1553 serial decode mode:
a Press the [Acquire] key.
b Press the Serial Decode softkey.
c In the Serial Decode Menu, press the Mode softkey.
d Use the Entry knob (or repeatedly press and release the Mode softkey) to select the
MIL-1553 serial decode mode.
2 Press Settings.
3 In the MIL-STD-1553 Settings Menu, use the Base softkey to select between
hexadecimal and binary display of the decoded data.
The base setting is used for the display of the remote terminal address and the data, in
both the decode line and in the Lister.
You can use the Zoom horizontal sweep mode for easier navigation of the acquired data.
"Probing MIL-STD 1553 Signals"
InfiniiVision Oscilloscopes User's Guide
on page 131.
"Setup for MIL-STD
"MIL-STD 1553