1. INTRODUCTION TO WAVEJET 300A REMOTE CONTROL INTERFACES
The WaveJet 300A supports remote control through USB, GPIB and TCP/IP. USB is supported as a standard
feature, GPIB and TCP/IP are available with optional interface boards.
The WJ-A-GPIB and WJ-A-LAN are optional interface boards exclusively for the WaveJe 300A series.
•
Insert the board into one of the option slots on the rear of the WaveJet 300A before use.
•
It can be inserted into either of the two option slots provided.
•
Once the board is inserted into the slot, the WaveJet main unit automatically recognizes it as a remote
interface, allowing you to select it.
•
The WJ-A-GPIB and the WJ-A-LAN can be inserted at the same time, but only one of them can be selected.
•
Two of the same kind of interface board must not be inserted at the same time (prohibited).
CAUTION!
To prevent damage to the instrument, turn instrument power OFF before inserting a WJ-A-GPIB or
WJ-A-LAN interface board.
ESD Sensitive
The WJ-A-GPIB and WJ-A-LAN interface boards are sensitive to electrostatic discharge
(ESD). To avoid causing damage to the boards, always follow anti-static procedures (wear
wrist strap, etc.) when handling the boards.
1.1 WJ-A-GPIB Hardware
Applicable standard
Usable unit
Power supply
Power consumption
Dimensions
Mass
Operating temperature range
Operating humidity range
Storage temperature range
1.2 WJ-A-LAN Hardware
Applicable standard
Usable unit
Power supply
Power consumption
Dimensions
Mass
Operating temperature range
Operating humidity range
Storage temperature range
917976 Rev A
WaveJet 300A Remote Control Manual
GPIB IEEE-488.1
WaveJet 3xx series firmware 3.xx or later
Supplied from WaveJet3xx option slot
Included in WaveJet3xx main unit
63 mm × 80 mm (excluding protrusion of connector terminal)
Approximately 70 g
0 to 40 °C
80% RH or lower (when operated within 0 to 40 °C)
-20 to +60 °C
TCP/IP 10BASE-T/100BASE-T
WaveJet 3xx series firmware 3.xx or later
Supplied from WaveJet3xx option slot
Included in WaveJet3xx main unit
63 mm × 80 mm (excluding protrusion of connector terminal)
Approximately 55 g
0 to 40 °C
80% RH or lower (when operated within 0 to 40 °C)
-20 to +60 °C