List Of Ieee 488.1 Interface Functions Subsets Implemented; Device Behavior When The Address Is Changed Outside The Range 0 To 30; When A User-Initiated Address Change Is Recognized By The Device; Device Setting At Power-On, Including Commands Which Modify The Power-On Settings - LeCroy WAVEJET 300A SERIES Manual

Oscilloscopes remote control
Hide thumbs Also See for WAVEJET 300A SERIES:
Table of Contents

Advertisement

WaveJet 300A Series

2.1 List of IEEE 488.1 interface functions subsets implemented

2.2 Device behavior when the address is changed outside the range 0 to 30

With GPIB, values that do not fall within the range of 0 to 30 cannot be selected.
TCP/IP has 4 fields, IPv4, for which values that do not fall within the range of 0 to 255 cannot be selected.

2.3 When a user-initiated address change is recognized by the device

GPIB immediately recognizes the address change.
TCP/IP recognizes the change when the system is rebooted after the change.

2.4 Device setting at power-on, including commands which modify the power-on settings

At power-on, device settings are restored to their values when the device was last powered off.
A command for changing the power-on settings is not installed.

2.5 Message exchange options

Size and behavior of input buffer

The size of the input buffer is 512 bytes.
Queries that return more than one <RESPONSE MESSAGE UNIT>
See the list of commands.

Queries that generate a response when parsed

See the list of commands.

Queries that generate a response when read

See the list of commands.

Commands that are coupled

See the list of commands.
8
Table 2-1. List of IEEE 488.1 interface functions
subsets implemented
Interface function
Subset
Source Handshake
SH1
Acceptor Handshake
AH1
Talker
T6,TE0
Listener
L4, LE0
Service Request
SR1
Remote Local
RL2
Parallel Poll
PP0
Device Clear
DC1
Device Trigger
DT1
Controller
C0
Electrical Interface
E2