LeCroy WAVEJET 300A SERIES Manual page 55

Oscilloscopes remote control
Hide thumbs Also See for WAVEJET 300A SERIES:
Table of Contents

Advertisement

Cursor
DESCRIPTION
COMMAND SYNTAX
QUERY SYNTAX
RESPONSE FORMAT
REMARKS
917976 Rev A
WaveJet 300A Remote Control Manual
VCUR
Command/Query
The VCUR command defines the position of the voltage
cursor. The VCUR? query returns the defined position.
VCUR <Cursor1>,<Cursor2>
<Cursor1>:=-3.97 to +3.95
<Cursor2>:=-3.97 to +3.95
<Cursor1> and <Cursor2> respectively represent
positions of voltage cursors Cursor1 and Cursor2 on the
screen, using the unit "div". For information on how
to round the value, see "Remarks". The bottom end of
the screen corresponds to -3.97, the center to 0, and
the top end to +3.95.
VCUR?
<Cursor1>, <Cursor2>
Current positions of the voltage cursors are identified
in <Cursor1> and <Cursor2>.
The format is <NR3 Numeric Response Data>.
The resolution of <Cursor1> and <Cursor2> is 0.02. If
other value is input, it is rounded to the number having
small absolute value.
VCUR command reports Device-specific error (DDE) when
CURM is not DV or DHDV.
55