LeCroy WAVEJET 300A SERIES Manual page 51

Oscilloscopes remote control
Hide thumbs Also See for WAVEJET 300A SERIES:
Table of Contents

Advertisement

Trigger
DESCRIPTION
COMMAND SYNTAX
QUERY SYNTAX
RESPONSE FORMAT
REMARKS
Others
DESCRIPTION
COMMAND SYNTAX
QUERY SYNTAX
RESPONSE FORMAT
917976 Rev A
WaveJet 300A Remote Control Manual
TSRC
Command/Query
The TSRC command defines the trigger signal source. The
trigger type determines the trigger signal source
setting. The TSRC? query returns the current trigger
signal source setting.
TSRC <source>
<source> := {CH1, CH2, CH3, CH4, EXT, EXT10, LINE} (for
WJ354/334/324/314)
<source> := {CH1, CH2, EXT, EXT10, LINE} (for
WJ352/332/322/312)
TSRC?
<source>
The current trigger signal setting is identified in
<source>.
When TTYP is not EDGE, TSRC LINE command is not
performed and reports device-specific error (DDE).
*TST?
Query
The *TST? query implements the internal self-test and
returns the response showing whether an error is
detected in the test.
*TST?
<status>
<status>:=+000000 No error occurred in the self test.
<status>:=-000001 CH1 ACQ MEMORY CHECK ERROR
<status>:=-000002 CH2 ACQ MEMORY CHECK ERROR
<status>:=-000003 CH3 ACQ MEMORY CHECK ERROR
<status>:=-000004 CH4 ACQ MEMORY CHECK ERROR
51

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents