WaveJet 300A Series
Data transfer
DESCRIPTION
COMMAND SYNTAX
QUERY SYNTAX
RESPONSE FORMAT
REMARKS
Data transfer
DESCRIPTION
COMMAND SYNTAX
QUERY SYNTAX
RESPONSE FORMAT
REMARKS
26
DTPOINTS
Command/Query
The DTPOINTS command defines the amount of waveform
data to be transferred. The
DTPOINTS? query returns the defined amount of waveform
data to be transferred.
DTPOINTS <points>
<points>:=1 to (waveform data length - DTSTART)
Any values exceeding this range are rounded to the
maximum or minimum permissible value, whichever is
nearest.
DTPOINTS?
<points>
The defined amount of waveform data to be transferred
is identified in <points>. The
format is <NR1 Numeric Response Data>.
(Transfer start address) + (Amount of data to be
transferred) > (Total amount of waveform data)
When DTPOINTS command causes the above case, the
transfer start address is rounded to a permissible
value.
DTSTART
Command/Query
The DTSTART command defines the transfer start address
for waveform data transfer.
The DTSTART? query returns the defined transfer start
address.
DTSTART <start>
<start>:=0 to (waveform data length - 1)
Any values exceeding this range are rounded to the
maximum or minimum permissible value, whichever is
nearest.
DTSTART?
<start>
The defined start address for waveform data transfer
is identified in <start>. The format is <NR1 Numeric
Response Data>.
(Transfer start address) + (Amount of data to be
transferred) > (Total amount of waveform data)
When DTSTART command causes the above case, the amount
of data to be transferred is rounded to a permissible
value.