Data transfer
DESCRIPTION
COMMAND SYNTAX
QUERY SYNTAX
RESPONSE FORMAT
REMARKS
Data transfer
DESCRIPTION
COMMAND SYNTAX
QUERY SYNTAX
RESPONSE FORMAT
REMARKS
917976 Rev A
WaveJet 300A Remote Control Manual
DTSTUP
Command/Query
The DTSTUP command writes setup data for the product.
The DTSTUP? query reads the setup data (the setup data
found when the query is issued) from the product.
DTSTUP<LF+EOI>#8<byte-length><binary-block>
DTSTUP command, unlike other commands, requires two
steps.
Step 1: Send DTSTUP command without parameter. Then the
scope becomes ready to receive setup data block.
Step 2: Send setup data block in
#8<byte-length><binary-block> format. The contents of
<byte_length> consists of an 8-digit ASCII string. This
string must represent the number of bytes to be entered
into <binary_block>. The byte length should always be
a sign-free, 8-digit integer, and zero must not be
deleted.
The data transferred by the DTSTUP? query should be
found in <binary_block>.
DTSTUP?
#8<byte-length><binary-block>
・#8 shows that the response message is described in
the format of <Definite Length Arbitrary Block Response
Data> and the content of the following <byte_length>
is a sign-free, 8-digit integer.
・The contents of <byte_length> represents the number
of bytes to be entered into <binary_block>. The byte
length should always be a sign-free, 8-digit integer
including zero.
・The contents of <binary_block> represent the
product's setup data in binary code.
This command does not support the multi-command.
DTWAVE?
Query
The DTWAVE? query reads the waveform data.
DTWAVE?
This command does not support the multi-command.