Remote Commands - LeCroy WAVEJET 300A SERIES Manual

Oscilloscopes remote control
Hide thumbs Also See for WAVEJET 300A SERIES:
Table of Contents

Advertisement

WaveJet 300A Series

6. REMOTE COMMANDS

This list of remote commands is alphabetized by short form.
Horizontal
DESCRIPTION
COMMAND SYNTAX
QUERY SYNTAX
RESPONSE FORMAT
REMARKS
Acquisition
DESCRIPTION
COMMAND SYNTAX
QUERY SYNTAX
RESPONSE FORMAT
Horizontal
DESCRIPTION
COMMAND SYNTAX
QUERY SYNTAX
RESPONSE FORMAT
REMARKS
Vertical
DESCRIPTION
COMMAND SYNTAX
QUERY SYNTAX
RESPONSE FORMAT
18
ACQ
Command/Query
The ACQ command selects the acquisition mode. The ACQ?
query returns the current selection of the acquisition
mode.
ACQ <mode>
<mode>:={NORMAL, PEAK, AVERAGE} (When XYDS is YT)
<mode>:={NORMAL, AVERAGE} (When XYDS is XYTRG)
<mode>:={NORMAL} (When XYDS is XY)
ACQ?
<mode>
ACQ AVERAGE sets ROLL OFF automatically.
ASET
Command
The ASET command provides the auto setup.
ASET
AVGCNT
Command/Query
The AVGCNT command sets the number of events to average.
AVGCNT? query returns the number set.
AVGCNT <number>
<number>:={2, 4, 8, 16, 32, 64, 128, 256}
AVGCNT?
<number>
AVGCNT command reports device-specific error (DDE) and
is not performed when ACQ is not AVERAGE.
BWL
Command/Query
The BWL command enables or disables the bandwidth limit
(low-pass filter) of the specified input channel. The
BWL? query returns the selected status.
<channel>: BWL <mode>
<channel>: = {C1, C2, C3, C4} (WJ354/334/324/314)
: = {C1, C2} (WJ352/332/322/312)
<mode>: = {OFF, 100M, 20M} (WJ354/334/352/332)
: = {OFF, 20M} (WJ324/314/322/312)
<channel>:BWL?
<mode>
The current bandwidth limit status (low-pass filter)
917976 Rev A

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents