LeCroy WAVEJET 300A SERIES Manual page 41

Oscilloscopes remote control
Hide thumbs Also See for WAVEJET 300A SERIES:
Table of Contents

Advertisement

Status
DESCRIPTION
COMMAND SYNTAX
QUERY SYNTAX
RESPONSE FORMAT
Status
DESCRIPTION
COMMAND SYNTAX
QUERY SYNTAX
RESPONSE FORMAT
Acquisition
DESCRIPTION
COMMAND SYNTAX
QUERY SYNTAX
RESPONSE FORMAT
917976 Rev A
WaveJet 300A Remote Control Manual
*SRE
Command/Query
The *SRE command defines the service request enable
(SRE) register. This command allows the user to specify
what SBT register summary bit is used to create the
service request (SRQ). If "1" is written to the
associated bit location, the summary message bit is
enabled. If "0" is written to the associated bit
location, the service request is no longer created by
the associated event. After the SRE register is cleared,
the SRQ interrupt is disabled. When the value is
converted into binary code, the *SRE? query
returns the value that represents the bit setting in the
SRE register.
*SRE <value>
<value>:=0 to 255
*SRE?
<value>
The current SRE register settings are identified in
<value> in <NR1 Numeric Response Data> format.
*STB?
Query
The *STB? query reads the status register (STB) defined
in IEEE 488.1 and the master summary status (MSS). The
response includes those values specified in bits 0 to
7 of the status byte register.
For details about structure of each bit, see Remarks.
The response to the *STB? query is the same as that for
series polling, except that bit 6 contains the MSS
summary message instead of the RQS message.
*STB?
<value>
The current STB register settings are identified in
<value> in <NR1 Numeric Response Data> format.
STOP
Command
STOP sets the trigger mode to "STOP". If the trigger mode
has already been "STOP", this command is ignored. STOP
acts same as "TRMD STOP".
STOP