LeCroy WAVEJET 300A SERIES Manual page 42

Oscilloscopes remote control
Hide thumbs Also See for WAVEJET 300A SERIES:
Table of Contents

Advertisement

WaveJet 300A Series
Trigger
DESCRIPTION
COMMAND SYNTAX
QUERY SYNTAX
RESPONSE FORMAT
REMARKS
Trigger
DESCRIPTION
COMMAND SYNTAX
QUERY SYNTAX
RESPONSE FORMAT
REMARKS
42
TCOUNT
Command/Query
The TCOUNT command defines the number of pulses for
pulse count trigger (TTYP=COUNT).
The TCOUNT? query returns the currently set number of
pulses for pulse count trigger.
TCOUNT <number>
<number> := 1 to 9999
TCOUNT?
<number>
TCOUNT command reports device-specific error (DDE) when
TTYP is not COUNT.
TCPL
Command/Query
The TCPL command defines the trigger coupling mode. The
TCPL? query returns the defined trigger coupling.
TCPL <tirg_coupling>
<trig_coupling>:={AC, DC, HF, LF}
TCPL?
<trig_coupling>
The current trigger coupling is identified in
<trig_coupling>.
When TSRC is LINE or TTYP is TV, TCPL command reports
device-specific error (DDE) and TCPL? query returns DC.