40
*SAV
Command
The *SAV command saves the front panel setup data of
the oscilloscope to nonvolatile internal memory.
*SAV <panel_setup>
<panel_setup>:0, 1 to 5
0 := do nothing.
SKLV
Command/Query
The SKLV command defines the measurement conditions of
automatic measurement SKEW specified with the MSEL
command.
The SKLV? query returns the defined measurement
conditions.
SKLV <level1>, <slope1>, <source2>, <level2>, <slope2>
<level1> is set in a range of 10 to 90 in <DECIMAL
NUMERIC PROGRAM DATA> format, assuming that P-P of the
waveform is 100%. A value beyond the setting range is
rounded to a permissible value.
<slope1>:={RISE, FALL}
<source2> :={CH1, CH2, CH3, CH4} (WJ354/334/324/314)
:={CH1, CH2} (WJ352/332/322/312)
<level2> is set in a range of 10 to 90 in <DECIMAL
NUMERIC PROGRAM DATA> format, assuming that P-P of the
waveform is 100%. A value beyond the setting range is
rounded to a permissible value.
<slope2>:={RISE, FALL}
SKLV?
<level1>, <slope1>, <source2>, <level2>, <slope2>
The parameter setting "Level1" of automatic
measurement SKEW is identified in <level1>.
The parameter setting "Slope1" of automatic
measurement SKEW is identified in <slope1>.
The parameter setting "Source2" of automatic
measurement SKEW is identified in <source2>.
The parameter setting "Level2" of automatic
measurement SKEW is identified in <level2>.
The parameter setting "Slope2" of automatic
measurement SKEW is identified in <slope2>.
If <from ch> or <to ch>is set to CH3 or CH4 in the model
WJ352/332/322/312, it is rounded to CH1.
SKLV command reports device-specific error (DDE) when
MSEL is not SKEW.
917976 Rev A