Other Remote Commands - LeCroy 9400A Manual

Digital oscilloscope
Table of Contents

Advertisement

7.6.6
Other
Remote Commands
1) CALIBRATE (CAL)
forces
the
9400A to do a calibration
of the
interpolator
and of both
input
channels
at
the
current
gain
and bandwidth
settings.
or
CALIBRATE (CAL)
,
<?
>
< ON >
*
< OFF >
*
allows
(OFF),
Section
itself.
automatic calibration (ON), disables automatic calibration
or queries
the state of the automatic
calibration.
See
9.4 for the conditions under which the 9400A calibrates
2) PROBE CAL (PC)
, < AC >
< DC >
, [ 0.000 V to 5.000 V ]
This command permits
generation
of a square signal (AC) with
period of 1.024 msec, or a continuous (DC) level at the specified
voltage at the probe test output.
When the 9400A returns to LOCAL, the default calibrator signal of 1
V (AC) is generated.
The 9400A sets the VALUE ADAPTED bit:
- if a positive out-of-range value is given.
3) AUTO STORE (AS) , ? >
<[0toS]
>
This
command allows the user to query the 9400A on its AUTOSTORE
state or setting of the AUTOSTORE mode:
O:
i:
2:
3:
4:
5:
AUTOSTORE off
automatic storage of CHANNEL 1 into MEMORY C
automatic storage of CHANNEL-1 into MEMORY-D
automatic storage of CHANNEL-2 into MEMORY-C
automatic storage of CHANNEL-2 into MEMORY-D
automatic storage of CHANNEL-1 into MEMORY-C
and of CHANNEL 2 into MEMORY-D
Remote
Operations
7-30