Chapter 4 Command Reference
[:SENSe] Subsystem
[:SENSe]:TINTerval Subtree (HP 53132A With S/N Prefix 3646
and Above)
This subtree controls the time interval (including Time Interval, Risetime,
Falltime, Duty Cycle, and Pulse Width functions) measuring capabilities of the
instrument . Refer to page 4-95, for Time Interval arming capabilities of the
HP 53131A (and the HP 53132A with a serial number prefix below 3646).
[:SENSe]:TINTerval:ARM:ESTART and :ESTOP Subtrees
(HP 53132A With S/N Prefix 3646 and Above)
These subtrees are used to synchronize the time interval start and stop arm with
events.
The following table defines the valid combinations of start/stop arming for Time
Interval only. HP recommends that SCPI programs issue arming commands in the
order shown in the following table ; that is, from left to right, beginning with
setting up ESTART and concluding by setting up ESTOP arming.