HP 53131A/132A 225 MHz Programming Manual page 61

225 mhz universal counter
Table of Contents

Advertisement

Keyword/Syntax
Parameter Form
:SYSTem
:COMMunicate
:SERial
:CONTrol
:DTR
IBFull | ON | LIMit
:TRANsmit
:BAUD
<numeric_value>
:PACE
XON | NONE
:PARity
[:TYPE]
EVEN | ODD | NONE
:ERRor?
:KEY
<numeric_value>
:LOG?
:VERSion?
:TRACe
:CATalog?
OFFSET, <numeric_value> [HZ
[:DATA]
| S | DEG]
SCALE, <numeric_value>
[:DATA]
OFFSET
[:DATA]?
SCALE
[:DATA]?
:TRIGger
:COUNt
:AUTO
<Boolean>
Chapter 2 Command Summary
Front Panel to SCPI Command Maps
Table 2-2. HP 53131A/132A SCPI Command Summary (Continued)
Programming Guide
Std/
Comments
New
Std
Subsystem. Collects the functions that are not
related to instrument performance.
Std
Subtree. Collects together configuration of
control/communication interfaces.
Std
Subtree. Controls the physical configuration of
the RS- 232C port.
Std
Subtree.
Std
Sets the usage of the DTR line of the RS-232
port.
Std
Subtree. Affects parameters associated with
transmission.
Std
Sets the baud rate.
Std
Sets the software pacing scheme.
Std
Subtree. Controls the parity of the channel.
Std
Sets the parity scheme.
Std
Query only. Queries the oldest error in the
Error Queue and removes the error from the
queue (first in, first out).
Std
Simulates the pressing of a front-panel key.
New
Query only. Returns a comma-separated list of
integers representing all of the entries in the
Key Queue.
Std
Query only. Returns the SCPI version number
with which the Counter complies.
Std
Subsystem.
Std
Query only. Returns list of intrinsic constants.
Std
Sets the offset value.
Std
Sets the scale value.
Std
Queries the offset value.
Std
Queries the scale value.
Std
Subsystem.
Std
Subtree.
New
Controls the number of measurements to be
made when :INIT[:IMM] is performed.
2-37

Advertisement

Table of Contents
loading

This manual is also suitable for:

53131a53132a

Table of Contents