Chapter 4 Command Reference
Introduction

:STATus Subsystem

The :STATus subsystem commands allow you to specify or examine the status of
the Operation Status Register group and the Questionable Data/Signal Register
group.

:STATus:OPERation Subtree

The :STATus:OPERation subtree commands allow you to examine the status of
the Counter monitored by the Operation Status Register group, shown in Figure 4-
2. The Operation Status Register group consists of a condition register, two
transition registers, an event register, and an enable register. The commands in this
subtree allow you to control and monitor these registers.
See the section titled "Operation Status Register Group and Questionable
Data/Signal Status Register Group" on page 3- 32 in Chapter 3 for a detailed
description of the Operation Status Register Group.
Figure 4-2. The Operation Status Register Group
:STATus:OPERation:CONDition?
Queries the status of the Operation Condition Status Register.
Bits are not cleared when read.
Query Response
Comments
The Operation Condition Status Register is cleared at power-on.
Programming Guide
Numeric data transferred as ASCII bytes in <NR1> format.
Range is 0 to 65,535.
The query response value is an integer formed by the
binary-weighting of the bits. The value of unused bits is zero.
4-107