Chapter 4 Command Reference
:CALCulate3 Subsystem
Related
Stats
Front-Panel
keys
:CALCulate3:AVERage:TYPE MAXimum | MINimum | SDEViation
| SCALar or MEAN
Selects which statistical result will appear:
Query Response
A sequence of ASCII-encoded bytes: MAX, MIN, SDEV, or MEAN
Comments
Related
Stats
Front-Panel
keys
:CALCulate3:DATA?
Queries the statistical result specified by :CALC3:AVER:TYPE.
Enable statistics (:CALC3:AVER[:STATe] ON) before attempting to query
results.
Query Response
4-22
in the :CALC3:DATA? response, and
on the front-panel display when :DISP[:WIND]:TEXT:FEED is set to
"CALC3".
*RST: MEAN
If :DISP[:WIND]:TEXT:FEED is "CALC3", then this command updates the
display immediately.
Result will be formatted according to :FORMat[:DATA] ASCii | REAL
setting.
When ASCii format is used, numeric data is transferred as ASCII bytes in
<NR3> format. The number of significant digits will range from 1 to 15,
depending on the measurement resolution. Only significant digits will be
returned.
If :CALC3:AVER[:STATe] is OFF, Not a Number 9.91E37 is returned and
error -221 is generated.
If no valid result exists, Not a Number 9.91E37 is returned and error -230 is
generated.