HP 53131A/132A 225 MHz Programming Manual page 269

225 mhz universal counter
Table of Contents

Advertisement

Chapter 4 Command Reference
Introduction
[:SENSe]:TINTerval:ARM:ESTOP[:LAYer[1]]:SOURce IMMediate
| TIMer | INTernal2
Sets or queries the stop arm for Time Interval measurements.
Query Response
A sequence of ASCII-encoded bytes: IMM or TIM or INT2
Comments
Related
Gate & ExtArm
Front-Panel
Keys
[:SENSe]:TINTerval:ARM:ESTOP[:LAYer[1]]:TIMer
<numeric_value> [S]
Sets or queries the time used to delay the stop arm for Time Interval
measurements.
<numeric_value>
For short delay: 100E-9 to 0.999,999,9 seconds
Range
For long delay: 1.00 0 to 10.000 seconds
<numeric_value>
For short delay: 100E-9 seconds
Resolution
For long delay: 1E -3 seconds
Query Response
Numeric data transferred as ASCII bytes in <NR1> format.
Comments
Related
Gate & ExtArm
Front-Panel
Keys
Programming Guide
*RST: IMMediate
The [:SENS]:TINT:ARM:ESTOP[:LAYer[1]]:SOUR command has no affect
on the following measurements: Risetime, Falltime, Duty Cycle, and Pulse
Width.
*RST: 100E-9 S
Only applies when [:SENS]:TINT:ARM:ESTOP[:LAYer[1]]:SOUR TIM is
selected.
The [:SENS]:TINT:ARM:ESTOP[:LAYer[1]]:TIM command has no affect
on the following measurements: Risetime, Falltime, Duty Cycle, and Pulse
Width.
4-103